咨询与建议

看过本文的还看了

相关文献

该作者的其他文献

文献详情 >Risks of inappropriate use of ... 收藏

Risks of inappropriate use of an algorithm favouring intrinsic conduction

作     者:Lo Savio, Armando Crea, Bruna Dattilo, Giuseppe Crea, Pasquale 

作者机构:Univ Hosp Messina Dept Clin & Expt Med Cardiol Unit Via Consolare Valeria I-98124 Messina Italy Univ Turin Sch Internal Med Turin Italy 

出 版 物:《PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY》 (整律和临床电生理学)

年 卷 期:2022年第45卷第12期

页      面:1345-1348页

核心收录:

学科分类:0831[工学-生物医学工程(可授工学、理学、医学学位)] 1002[医学-临床医学] 100201[医学-内科学(含:心血管病、血液病、呼吸系病、消化系病、内分泌与代谢病、肾病、风湿病、传染病)] 10[医学] 

主  题:algorithm pacemaker RYTHMIQ (R) ventricular pacing 

摘      要:An 80-year-old woman received a dual chamber pacemaker (Boston Scientific Accolade MRI DR) for pre-syncopal episodes associated with transient II-degree atrioventricular block type 1 and 2:1, recorded in 24-h Holter monitoring. Due to residual AV conduction with I-degree AV block, the pacemaker was set with the RYTHMIQ (R) algorithm, in order to reduce inappropriate ventricular pacing. A month later the patient started to complain of severe asthenia and bradycardia (46-48 bpm). Telemetry-supported pacemaker control revealed III-degree AV block with junctional escape rhythm, unmasking missed switch of RYTHMIQ (R) algorithm.

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分