Considerations used to select minicomputer software and hardware appropriate for radiotherapy patients’ records storage and retrieval are discussed. (AIP)
Considerations used to select minicomputer software and hardware appropriate for radiotherapy patients’ records storage and retrieval are discussed. (AIP)
A computer interface for the Clinac-18 linear accelerator has been developed, using a standard CAMAC interface plus buffer amplifiers to isolate the CAMAC from the accelerator electronics. Buffer amplifiers are employ...
详细信息
A computer interface for the Clinac-18 linear accelerator has been developed, using a standard CAMAC interface plus buffer amplifiers to isolate the CAMAC from the accelerator electronics. Buffer amplifiers are employed because direct connection of the CAMAC system to the accelerator was found to affect accelerator operation. The total interface accommodates the four gantry position analog signals and thirteen digital signals describing all available treatment options. The interface also allows the computer to inhibit beam operation.
A series of computer programs have been written to analyze a scan of the AAPM CT performance phantom resolution insert with little operator assistance. A representative analysis is done on a scan from an AS&E scan...
详细信息
A series of computer programs have been written to analyze a scan of the AAPM CT performance phantom resolution insert with little operator assistance. A representative analysis is done on a scan from an AS&E scanner as an example. Its point spread function, edge response function and modulation transfer function are presented.