Direct analysis in real time mass spectrometry (DART-MS) method has been developed using the Quality by Design (QbD) *** this presented approach, method performance was mathematically described as a composite desirabi...
详细信息
Direct analysis in real time mass spectrometry (DART-MS) method has been developed using the Quality by Design (QbD) *** this presented approach, method performance was mathematically described as a composite desirability function of the critical quality attributes (CQAs).Two terms of method validation, including analytical repeatability and method robustness, were carried out at an operating work *** presented strategy provided a tutorial to the development of a robust QbD compliant mass spectrometry method for industrial quality control.